< No response from wrapper (402)

|
Error content |
There was no response with respect to the inquiry conducted by the main board to the wrapper board P-1105 [A200] about the state. |
|
Detail |
This is a communication error that has occurred after performing power-on processing. The sub-error includes the communication phase number. |
Solution |
Press [STOP] to release the error. Press [REEXEC] to check the connection with the wrapper again. |
User’s action |
Turn off the power supply and then turn on the power supply again. |
Service person’s action |
|
Related part |
① P-1105 [A200] XJ26 ② DC 24V supply to the power supply (XJ12) of P-1105 [A200] (DC 24V is normally supplied when the LED of P-1105 [A200] is lit.) ③ RS485 communication cable between the main board (P-1100, P-1163, P-1000) [A10] and P-1105 [A100, A200, A300] The following RS485 communication has not been established: Between P-1163 [A10] (XJ4) and P-1105 [A100] (XJ26) Between P-1105 [A100] (XJ27) and P-1105 [A300] (XJ26) Between P-1105 [A300] (XJ27) and P-1105 [A200] (XJ27) |
|
Remarks |
Pricing is only possible after the [STOP] button is pressed. |